LEMON Manuals: Even more car manuals for everyone: 1960-2025
Home >> Buick >> 2016 >> Envision >> Repair and Diagnosis >> External Pages >> Different car >> Section 199 (Object Detection System) >> Repair Instructions >> Front View Camera Module Calibration >> Calibration
April 5, 2026: LEMON Manuals is launched! Read the announcement.

Front View Camera Module Calibration: Calibration

WARNING: This page is about a different car, the 2016 Buick LaCrosse. However, it is still accessible from the selected car via links, so may be relevant.

For North American SPS users, Calibration will NOT start automatically  and will need to be initiated with GDS2. For European SPS users, Calibration will start automatically  and will not need to be initiated with GDS2.

The calibration process for the front view camera system is necessary when a front view camera module is replaced by a new one. This process shall not be required when only replacing the windshield and the front view camera module has been mounted again properly. This process shall be completed within 3-5 minutes when the correct driving conditions are met. If conditions are not met correct, the front view camera module shall continuously run the service point calibration until successfully completed. This process shall work across ignition cycles and shall not be required to be re-initialized at start up. If the Lane Departure Warning Switch is pressed during calibration, the indictor will flash momentarily and then stay out - this is normal operation.

To calibrate the front camera, operate the vehicle in the following conditions until the calibration is complete:

Calibration Procedure (Required for North American SPS users)

  1. Apply the parking brake.
  2. Place the transmission in Park (if applicable)
  3. Install scan tool
  4. Navigate to the module diagnostics menu
  5. Select Frontview Camera Module
  6. Select Configuration/Reset Functions menu Item
  7. Select the Frontview Camera Learn procedure and follow the directions displayed on the screen.
  8. Drive the vehicle to complete calibration.

Once the procedure is complete, the amber indicator will turn off. Shortly after the green ready to assist light should turn on as long as all conditions are met for normal operation. The system is then ready to assist.