Component Testing
Skewed Sensor Test
- Ignition OFF, remove the vacuum source from the sensor.
- Ignition ON, observe and record the scan tool sensor pressure parameter. This is the first sensor reading.
- With the J 23738-A or J 35555 , apply 5 in Hg (17 kPa) of vacuum to the sensor. Observe and record the scan tool sensor pressure parameter. This is the second sensor reading.
- Subtract the second sensor reading from the first sensor reading. Verify that the vacuum decrease is within 1 in Hg (4 kPa) of the applied vacuum.
- If the vacuum decrease is not within the specified range, replace the sensor.
- With the J 23738-A or J 35555 , apply 10 in Hg (34 kPa) of vacuum to the sensor. Observe and record the scan tool sensor pressure parameter. This is the third sensor reading.
- Subtract the third sensor reading from the first sensor reading. Verify that the vacuum decrease is within 1 in Hg (4 kPa) of the applied vacuum.
- If the vacuum decrease is not within the specified range, replace the sensor.
Erratic Signal Test
- Ignition OFF, remove the sensor.
- Install a 3A fused jumper wire between the 5-volt reference circuit terminal C and the corresponding terminal of the sensor.
- Install a jumper wire between the low reference circuit terminal A of the sensor and ground.
- Install a jumper wire at terminal B of the sensor.
- Connect a DMM between the jumper wire from terminal B of the sensor and ground.
- Ignition ON, with the J 23738-A
or J 35555
, slowly apply vacuum to the sensor while observing the voltage on the DMM. The voltage should vary between 0-5.2 volts without any spikes or dropouts.
- If the voltage reading is erratic, replace the sensor.